Mettler Toledo® XP2U (11122430) Ultra-Micro Balance 2.1 g x 0.1 µg |
|
METTLER TOLEDO XP2U offer unsurpassed accuracy up to a tenth of a microgram. Even sample sizes below 1 mg still meet strict process tolerances as well as comply with international regulatory standards. more ... |
Laboratory Equipment >
Analytical Balances >
Ultra Micro - 0.0001 mg
|
Specifications Chart
|
Have Questions? Our Staff will be glad to help |
Capacity x Readability
2.1 gx 0.0001 mg
Pan Size
16 mm (Diameter)
|
|
|
|
|
|
|
Features
XP2U Ultra Micro Balance
Ensure maximum sample yield for valuable substances. Keeping sample weights as small as possible means tremendous cost savings when handling precious, degradable or toxic substances.
METTLER TOLEDO XP2U offer unsurpassed accuracy up to a tenth of a microgram. Even sample sizes below 1 mg still meet strict process tolerances as well as comply with international regulatory standards.
The XP6U / XP2U and XP6 Micro Balances offer:
- • Optimum measurement performance in seconds
- • Comprehensive QM-functions that help to fulfil regulatory requirements
- • Intuitive and easy operation
METTLER TOLEDO XP2U Micro Balance and XP6 Ultra Micro Balance are designed to boost efficiency and reliability as well as support network compatibility.
- Reliable results - With a readability of 0.1 ìg and an unparalleled capacity of 6 g the Ultra Micro Balance sets new standards for weighing down to a tenth of a microgram.
- Save costs - Excellent repeatability helps to reduce sample weights and the short stabilization time of only 8 seconds maximizes sample throughput. All in all METTLER TOLEDO Micro Balances cut your costs significantly.
- Easy operation - With its clearly structured and customizable color touchscreen and SmartSens infrared sensors for hands-free operation, the terminal ensures simple, fast and error-free use
- Audits passed easily - Inbuilt safety features such as password protection and minimum weight control make the XP6U / XP2U and XP6 the ideal instrument for regulated environments
Standard Equipment
- ProFACT: fully automatic time- and temperature-controlled internal adjustment and linearization
- Easy to clean, motor-driven draft shield
- SmartSens: two IR sensors for hands-free operation
- In-built RS232C, two auxiliary interfaces for connecting a keyboard or additional IR sensor for hands-free operation
- Slot for a second optional interface such as LocalCAN, Ethernet, RS232, USB, MiniMettler, Bluetooth, PS/2
Colored touchscreen display
- 7 different pre-programmed applications
- 8 configurable user accounts with password protection
- 4 definable info fields for user and sample identification
- SmartTrac: graphic weighing-in aid, to track capacity and weighing tolerances
- 3 definable minimum weights with MinWeigh warning if the minimum weight requirement is violated • Up to 10 shortcut keys for special functions
Ergonomic weighing - Height adjustable weighing table for ergonomic weighing
Electrostatic Discharging - System controlled U-Ionizer for efficient discharging
Excellent Connectivity - 7 freely definable optional interfaces
Intuitive User Interface Customizable colored touchscreen display
With up to 61 million resolution points, XP Micro and Ultra Microbalances are ideal for accurately measuring the smallest of samples
- Benefit from GWP® Excellence Only GWP® Excellence balance firmware provides you with integrated routine testing procedures for guaranteed compliance with your SOP
- Minimum Weight The MinWeigh function helps to comply with the minimum sample weight required by USP. Auditors view the MinWeigh certificate as a QM document
- Easy Operation with LabX The user can access predefined weighing applications from the display and work through them step by step using the intuitive user prompts.
- Optical Guide Tolerances can be set individually. The optical guide helps to achieve the smallest possible sample weight.
Weight:
22.00 lb
|
|
Specifications
Model |
XP2U 11122430 |
Capacity x Readability |
2.1 g x 0.0001 mg |
Repeatability at nominal load |
0.00025 mg |
Repeatability at low load (measured at) |
0.0002 mg (0.2 g) |
Linearity |
0.001 mg |
Eccentric load deviation (test load)1) |
0.0025 mg (1 g) |
Sensitivity offset |
1.5 x 10–5 |
Sensitivity temperature drift2) |
0.0001%/°C |
Sensitivity stability3) |
0.0001%/a |
Typical values4) |
Repeatability (sd) |
0.00015 mg+2.5 x (10–8)·R_gr |
Differential linearity deviation (sd) |
8 x (10-14) gER_nt |
Differential eccentric load deviation (sd) |
8 x (10–7)·R_nt |
Sensitivity offset (sd)2) |
3 x (10–6)·R_nt |
Minimum weight* (according to USP) |
0.45 mg+7.5 x (10–5)·R_gr |
Minimum weight* (@ U = 1%, 2 sd) |
0.03 mg+5 x (10–6)·R_gr |
Settling time |
< 10 s |
pan size Ø |
16 mm |
1) According to OIML R76
2) In the temperature range 10…30°C
3) Stability of sensitivity as from first installation with FACT
4) Can be used for the estimation of uncertainty
sd: standard deviation Rgr: gross weight Rnt: net weight (sample weight) a: year (annum)
* Repeatability and minimum weight can be improved and affected by the following measures:
– choice of suitable weighing parameters, – moving to better location, – using smaller tare containers
|
|
|
* within the 48 continental US States
|
|
|
|